Contamination Beware: Cleaning 101 for Fibre Optics


Posted by Camilla Andersson on October 24, 2012

 Grasping a few basic cleaning techniques can save you time and money.


Over the past decade, the use of fibre optics in the medical industry has increased steadily. From extensive medical applications, including light therapy, arthroscopic surgery, x-ray imaging, ophthalmic lasers, and lab and clinical diagnostics, to simple communication processes, such as transmitting patient information, fibre optics help medical professionals conduct procedures that were previously not possible.

A fibre optic cleaning stick being used to clean an LC connector located in a mating sleeve. The stick is engineered to reach down into the mating sleeve and clean the endface of a fibre optic termini located within. Photo courtesy of MicroCare, Sticklers.

Maintenance of the connector endface is imperative for the fibre optic instrument to work correctly. It is an exacting job: the microscopic termini that create a connection must meet at an exact place, and the glass cores need to be perfectly aligned in order to have signal transmitted to and from the device. In addition, the termini endfaces must be completely clean of contaminants to ensure minimal signal loss. One of the most basic and important procedures for the maintenance of fibre optic systems is to clean the fibre optic termini.

Contamination challenges


Sidebar: Do’s and Don’ts when Cleaning Fibre Optics

 

Cleanliness is crucial to the performance of fibre optic instruments. Any contamination on the termini endface can cause failure of the component or the system as a whole. Even microscopic particles on the endfaces can cause a variety of problems for optical connections. One of the biggest challenges of contamination is that it cannot be seen with the naked eye, and one must typically use a specialised 200x or 400x inspection scope to determine the cleanliness of the endface. A close inspection of the connector is critical to identify that particles and residue are completely eliminated, thereby ensuring that connections work to their full potential.

A variety of serious, undesired outcomes could occur if a contaminated fibre optic connection is used, from deteriorated performance levels to ruined instruments. A dirty fibre optic endface will significantly degrade signal transmission and can result in blocking the signal all together. Even if a particle is only situated on the ferrule or the edge of the endface, it can cause an air gap or misalignment in the termini between the cores. This can result in back reflections, instability in the laser system, signal attenuation or even a system shutdown. Another potential issue is a scratched surface as a result of dust particles trapped between two termini endfaces. In addition, some fibre optic instruments, such as those using high-power Class IV lasers, generate a significant amount of heat, which, when in contact with contaminants, can spark a violent reaction or fire.

Inspection scope photos at 400X magnification of a pristine clean termini endface. Note the actual glass fibre is enclosed in the small white dot seen in the clean and particulate views.
Inspection scope photos at 400X magnification of a termini endface contaminated with finger oil. The fibre is completely obscured. Note the actual glass fibre is enclosed in the small white dot seen in the clean and particulate views.

One of the biggest challenges related to cleaning fibre optic termini is finding a process that works. Improvising a cleaning process will almost certainly lead to failure, as previously outlined. The best advice is to inspect, clean and inspect again. Repeat this process until you are absolutely sure the instrument is clear of all contaminants. Spending the time to clean it right the first time saves time and money in the end.

Cleaning methods 101
Materials used to clean the endfaces must be pristine; otherwise you could make it worse by adding contamination. It may be intuitive to wipe the endface on your gown or a cloth, but under a typical fibre optic inspection scope, those items carry a large variety of contaminants that could soil the connector. Even touching the termini with your finger will cause it to be significantly dirtied with skin oil. To avoid further contamination, make sure to use a product that has been specifically manufactured for cleaning fibre optics. Always wash your hands prior to using cleaning materials to minimise the transfer of skin oil onto the cleaning surfaces.

There are two primary methods available to properly clean fibre optic termini: a specialty wipe for male connectors and a swab for female connectors. Particulates, oils and salts are the three basic types of contaminants found on the endfaces, all of which require their own cleaning methods. Particulates are solids that are held on the endface by a static attraction. The best way to clean these types of contaminants is by dissipating the static charge that both attracts and holds them in place, which usually can be done with a specialty cleaning fluid. Cleaning fluids also are used to dissolve oils found on the fibre. Salts, on the other hand, are not necessarily fully removed by cleaning fluids alone. While cleaning fluids may quickly rinse away the oils, they tend to leave salt remains behind in the form of a white residue that can be very difficult to remove. Mechanical action that a wipe or a swab provides is usually combined with a cleaning fluid to fully eliminate oil and salt left on the termini endface.

Buyer beware, it’s fair to say that some cleaning products on the market can leave the endface dirtier than when you started. To avoid this situation, look for products that offer engineered solutions formulated to rid the termini endface of particulates, oils and salts. Avoid using aqueous (water-based) cleaning solutions or pure isopropyl alcohol (IPA), as violent reactions could potentially occur. Aqueous products are slow to dry and can leave moisture on the endface. In cold ambient temperatures, the moisture may actually freeze on the endface and alignment sleeve. If the moisture is not completely removed before the fibre is connected in the sleeve, the laser-energised fibre will instantly vaporise the remaining liquid into a gas, causing an explosion through sudden expansion of the vapours. As with water-based cleaners, IPA may explode or catch on fire when left on a highly energised fibre endface. Adding insult to injury, IPA frequently leaves a hazy film behind when it dries. What you should look for is a fast-drying, high-purity fluid engineered specifically for cleaning fibre optics.

Inspection scope photos at 400X magnification of a termini endface contaminated with particulate. Note the actual glass fibre is enclosed in the small white dot seen in the clean and particulate views.Photos courtesy of MicroCare, Sticklers

High-purity cleaning fluids can and should be used with both wipe and swab applications. However, beware of presaturated cleaning materials. Presaturated wipes and swabs often contain microscopic contaminants drawn from the plastic packaging, which will transfer to the endface during the cleaning process and result in further contamination problems. Instead, carefully apply a small amount of high-purity cleaning fluid on the corner of a dry wipe or the tip of the swab and then apply to the fibre optic termini. A well-engineered cleaning fluid will not only dissolve oils found on the endface, but will help to eliminate the electrostatic charge generated when the applicator is pulled out of its packaging. Be sure to not touch the area of the wipe or swab you will be using with your finger or your clothing. Should you touch this area or drop it on the ground, discard the wipe or swab and start over. Once the cleaning process is complete, discard the wipe or swab and inspect the endface to make sure all contaminants have been eliminated.

Bottom line: Performance-critical cleaning
As fibre optics usage continues to grow in the medical industry, it is imperative to clean the termini endfaces properly the first time around with the right products. The consequences of an unclean fibre optic can be detrimental to your business operations. Medical professionals can’t afford to have a blocked connection, a fire, or a complete system shutdown when there is a patient on the table. Use the correct cleaning solutions and closely inspect the endface to avoid these costly mistakes. Investing the time, energy and money into the cleaning process at the beginning will end up saving you in the end.

 

 

Jay Tourigny is
Vice President of Operations at MicroCare Medical



Find more content on:
No votes yet


Login or register to post comments

Comment viewing options

Select your preferred way to display the comments and click "Save settings" to activate your changes.

The fiber optics are just

The fiber optics are just great. We need to get ourselves acquainted with this technology. - YOR Health Products

fe Impressive article and

fe Impressive article and lucky are the readers of this site for every information provided here is reliable. pdf to flash converter

Reply

The fiber optics were in use in the ancient centre. With the advancement in the technology, medical filed has increased the use of it in the areas of light therapy, arthroscopic surgery, x-ray imaging. The job of the medical field authorities have become much more easier with the implementation of the fiber optics.
furniture removal

Good

A well-engineered cleansing watery devise negative merely evanesce lubricates initiate on the endface, yet devise auxiliary to eradicate the electrostatic mandate produced during the applicator is pulled away of its packaging.
mba thesis