Mechanical Testing of Medical Implants


Posted in Testing Equipment by Camilla Andersson on May 11, 2012

Implants and the materials used for their production undergo extensive tests, which include the evaluation of the static and dynamic mechanical properties of the raw materials, components and systems. Many of these tests are internationally standardised. Other testing procedures are specific to a given application. In all cases, advanced testing instruments and clamping fixtures, along with flexible, programmable testing software, help in the determination and analysis of the relevant data.


Manufacturers of medical implants must demonstrate that they have carefully considered the risks of component or system failure and satisfactorily mitigated against them. To this end, such products undergo a variety of mechanical tests prior to their approval, from simple tensile tests on material specimens, through combined compression, torsion or flexure tests, up to the simulation of pressure pulsation on complete implant systems under defined ambient conditions. The examples below illustrate the wide range of testing applications.

Knee replacements: Fatigue behaviour of tibial trays
Fatigue fracture of knee tibial trays has been one of the most commonly reported failure mechanisms in Total Knee Replacements (TKR). It is caused by loss of underlying bone support underneath the implant, which renders the latter mechanically unstable. Cyclic loads imparted by walking cause fatigue cracks, ultimately leading to catastrophic failure. The ISO 14879-1 standard1 specifies a set of test parameters for determining the fatigue properties of different tibial tray designs.

In practice, servo-hydraulic test systems such as Instron's model 8870 have been found eminently suitable both for determining fundamental material properties such as the resistance to fatigue crack propagation, and testing of the entire tibial tray. One half of the tibial tray is secured in a clamping fixture, simulating a fully supported condyle. The other, unsupported condyle is then subjected to physio¬logically representative loading.

 Figure 1. Compact and space-saving servo-hydraulic test system, prepared for testing the fatigue properties of tibial trays for total knee replacements.

A fatigue testing system used frequently for this type of test is Instron's model 8872, a compact table-top test instrument (Figure 1). With a nominal load capacity of 25 kN it is particularly suited for static and dynamic testing of materials and components for biomedical applications. Its adjustable crosshead with integral servo-hydraulic actuator, which accommodates the load cell, and a corrosion-resistant base with T-slot table and drain channel, make the instrument eminently suitable for testing in a saline bath, as well as a wide range of other tests. The Console software provides extensive features for PC control of the testing system, including signal generation, calibration, limit setup and status monitoring. In addition, the Wavematrix block programming software can be used for cyclic tests.

Typically, when performing such measurements under dynamic loads, acceleration forces are acting on parts of the system, which are superimposed on the actual forces applied to the specimen. Therefore, the load cells will also record forces resulting from their own movement and that of the clamping fixtures attached to them. Dynacell load cells (with load capacities ranging from 250 N to 2500 kN) reduce these effects even at high test frequencies. In conjunction with the associated 8800 Digital Controller, they enable the automatic compensation of acceleration forces as well as inertial errors such as those caused by hydrodynamic effects that occur when testing in a fluid bath.

Spinal columns: Static, torsion and fatigue testing
In spinal constructs, high loading may lead to catastrophic failure. Hence, static testing must be performed to evaluate the compressive, tensile and torsional loads leading to the fracture of the spinal construct. In addition, service life testing is also critical, as fatigue fracture is more common in actual practice than catastrophic failure. Loading typically is applied with a constant-amplitude, load-controlled sinusoidal waveform, with tests running in excess of five million cycles. The ASTM F1717 standard2 specifies test methods for both static and fatigue testing. Testing is usually done with blocks of ultrahigh molecular weight polyethylene (PE-UHMW) rather than vertebrae, to eliminate vertebra-induced variance.

 Figure 2. Electro-dynamic testing systems offer exceptional response and high accuracy across a wide frequency range for fatigue testing of spinal constructs.

 

Universal testing systems are generally well suited for static tensile and compressive testing. Standard test software packages (e.g. Instron's Bluehill 3) can be used to record load-displacement curves and perform the calculations specified in the standard. Table-top torsion testers such as the 55MT MicroTorsion systems offered by Instron (designed for multiple revolutions) are suitable for application of torsional loads.

If an angle of rotation of ±135° is sufficient, the model 8874 fatigue testing system is an efficient solution for handling both axial and torsional fatigue testing. The biaxial, table-top servo-hydraulic instru¬ment features a dual-column frame and a combined dynamic actuator for axial and torsional loading in the upper crosshead. Once again, the 8800 digital electronic controller and the Dynacell load cell in conjunction with the Console and/or WaveMatrix software are ideal complements to the system.

If you need to test the durability of spinal constructs and require exceptional response and accuracy across a wide frequency range, Instron's all-electric test instruments (such as the ElectroPuls E1000, Figure 2), combined with the WaveMatrix software, provide the necessary capabilities. These advanced instrument feature a digital control system for easy loop tuning based on specimen stiffness.

The system is fed from a single-phase ac power supply and requires no other utilities for basic operation (such as pneumatic air, hydraulics or water).

Stents: Efficient fatigue testing of materials and structures

 Figure 3. A multi-specimen fixture assists manufacturers of cardio¬vascular implants in the evaluation of the fatigue strength of nickel titanium (Nitinol) and stainless steel materials and structures.

 

Traditionally, testing of stents and stent grafts to standards such as ASTM F2477–07.3 involves subjecting complete devices to pressure pulsation simulating blood flow for a prespecified number of cycles. To enable a representative sample of specimens to be evaluated and to reduce overall test time, multiple specimens must be tested on a single system.

To cater to this requirement, Instron has developed a multispecimen fixture for the ElectroPuls E3000 test instrument (Figure 3). Each specimen station features a fatigue-rated load cell, precision alignment and grips that are unique to the material or structure undergoing test. The entire assembly is submerged in a temperature-controlled bath to test the specimens in vitro. The test is controlled by the WaveMatrix software, which features integrated temperature control of the fluid bath, a live display of each load cell reading and trend monitoring of forces to determine specimen fracture.
 

Wide-ranging experience
The solutions described above only represent a cross-section of a diverse spectrum of standardised and application-specific testing solutions developed by Instron in close partnership with users from the medical device industry. A more comprehensive overview can be found on the website http://go.instron.com/biomedtesting. In all cases, the test system manufacturer's application specialists will be prepared to configure, in close cooperation with manufacturers and users, testing systems tailored to specific testing requirements, and to develop unconventional solutions for challenging testing tasks.

References
1. ISO 14879-1: Determination of endurance properties of knee tibial trays
2. ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model
3. ASTM F2477–07: Standard Test Methods for in vitro Pulsatile Durability Testing of Vascular Stents

 



Find more content on:
No votes yet


Login or register to post comments

Comment viewing options

Select your preferred way to display the comments and click "Save settings" to activate your changes.

The post is written in very a

The post is written in very a good manner and it contains many useful information for me.
get him to propose and move the relationship forward

Clinical Laboratory Science

Clinical Laboratory Science is the second name of Medical Technology (MT), this blog is helpful for me because I am a writer at essay help online at speedyEssay writing an assignment on What is a medical technologist, and what do medical technologists do.

I went over this website

I went over this website and I believe you have a lot of wonderful information, saved to my bookmarks e zigarette liquid

One half of the tibial tray

One half of the tibial tray is secured in a clamping fixture, simulating a fully supported condyle. The other, unsupported condyle is then subjected to physio¬logically representative loading.
http://rinatlazar.bloger.co.il/129096/

Once again, the 8800 digital

Once again, the 8800 digital electronic controller and the Dynacell load cell in conjunction with the Console and/or WaveMatrix software are ideal complements to the system.
serrurier paris 7

website is great

Hi top information in your website and content is updated also. good fap turbo forex trade software 2014

Medical device technology to

Medical device technology to the marketplace is a high‐risk venture. That is why we strive to provide our clients with absolute testing certainty by providing accurate and reliable results for our clients’ medical device testing needs, from small, start-up ventures to large, blue chip manufacturers.Assignment Plus

In addition, service life

In addition, service life testing is also critical, as fatigue fracture is more common in actual practice than catastrophic failure.
ask.fm hack

The Console software provides

The Console software provides extensive features for PC control of the testing system, including signal generation, calibration, limit setup and status monitoring. In addition, the Wavematrix block programming software can be used for cyclic tests.
serrurier paris

nice

Hello,i really like your website design... its great... petroleum engineering schools with How to Become an Electrician

This is a best currently

This is a best currently emerging technologies, which contains some of the most prominent ongoing developments, advances, and innovations in various fields of modern technology.Expert assignment uk

The plant testing will bear

The plant testing will bear new grounds. If successful, then it would be great. - Kris Krohn

Technology adoption in

Technology adoption in medical field is good thing for next generation and health. Get custom business logo design for your company.

Medical technology having

Medical technology having developed into new emerging trends. Here are superior papers reviews for your help for your college and study work.

We should care more about

We should care more about medical implants. It is very important because it can save people’s lives! The testing standards are very good. But what I am really concerned about is the corruption. It is very bad side of business. We should monitor the global changes in the medical manufactures too in my opinion. If you have any problem with windows issues, do check the link, it is very helping…
Virus Removal

Recherche dans Video m’a

Recherche dans Video m’a conduit ici, je viens de trouver ce genre de lectures satisfaisante que je cherchais for. pdf to flash converter